TM 9-4120-378-14
REFRIGATION SYSTEM SERVICING - (EVACUATION) - Continued
5.8
Remarks
Action
Location/Item
e. Open charging manifold valves.
f . Run the vacuum pump until approximately a 300
micron vacuum is reached.
NOTE
Inability to reach 300 microns may indi-
cate either a leak or a problem with the
pump.
Close manifold valves and check compound gage.
g.
Record reading. Let unit sit for 1 hour. Observe
compound gage reading.
if the system holds the vacuum without change of
h.
pressure, proceed to Step j.
i.
If the vacuum cannot be held for 1 hour, one of the
following reasons may account for the problem:
(1)
Presence of water vapor in the system. Con-
tinued pumping will correct this condition.
(2)
vacuum with dry nitrogen and retest for leaks.
(3)
Internal leakage of vacuum pump. Test the
pump by connecting a vacuum gage directly
to the vacuum pump intake and continue to
pump. If pump still fails to reach approximately
300 microns, the pump is faulty.
j . Close both unit service valves.
k.
Close charging manifold valves.
l.
Stop vacuum pump.
m.
Disconnect pump from center hose connection.
n.
Charge system with refrigerant-22.
5-23