TM 9-4910-700-14 & P
6.
Connect the loop and adjust
3.
Adjust the repetition rate
R14 on the test unit for an
using the electronic counter
exact indication of 5000RPM.
to 8.333PPS (1000RPM). Re-
connect the loop to the out-
7.
Repeat step 2 and adjust the
repetition rate for 20.833PPS
Reconnect the
(2500RPM).
4.
Adjust R16 for an exact
The meter should read
loop .
1000RPM indication on the
between 2350 and 2650 RPM.
meter under test.
5.
Repeat step 2 and adjust
the repetition rate to
41.66PPS (5000RPM).
4-5. TROUBLESHOOTING. Troubleshooting
4-6. REPAIR, Repair procedures for the
may be best accomplished by use of the
test set are limited to replacement
following procedure.
of defective components.
a.
Replacement of Board Mounted
a.
Table 4-3 lists some symptoms
Electronlic Components. W hen re-
and probable defects for the
placing parts on circuit boards,
it is recommended that a solder-
and follow the indicated fault
ing iron of 40 watts or less be
isolation procedures.
used. After removing the com-
ponent, clear the solder out of
b.
the holes in the circuit board
Troubleshooting. Table 4-4 lists
before re-inserting. Remove flux
residue with Freon TMC or Tri-
the dwell meter and points resis-
tance circuit.
b.
Replacement of Chassis Mounted
Components. C hassis mounted comp -
orients should be replaced using
proper mechanical assembly pract-
ices . Be sure to retain all mount-
ing hardware when replacing parts.
Refer to chapter 5 for parts
listing.