TM 9-4935-481-14-1
C2
Table 3-11. DMS-D Performance Test-Continued
Table 3-11. DMS-D Performance Test-Continued m
STEP
UNIT
PROCEDURE
CORRECTIVE ACTION
STEP
UNIT
PROCEDURE
CORRECTIVE ACTION
SEQUENCE
Reading
SEQUENCE
Reading
231
AUTO 13
+0.3218 to +0.3258 Vdc
244
AUTO 1
0.4216 to 0.4268 Vdc
232
AUTO 14
+0.3218 to +0.3258 Vdc
Replace card A1. Check A1 pins 62, 63 and 72.
233
AUTO 15
+0.3218 to +0.3258 Vdc
245
AUTO 2
+4.201 to +4.226 Vdc
234
AUTO 16
+0.3218 to +0.3258 Vdc
246
AUTO 5
+4.201 to +4.226 Vdc
235
AUTO 4
+0.3285 to +0.3529 Vdc
247
AUTO
6
+4.201
to
+4.226
Vdc
Corrective action for steps 220 through 235 is:
Replace
248
AUTO
8
+4.201
to
+4.226
Vdc
card A1. Check A1 pins 62, 63 and 72.
249
AUTO
9
+4.201
to
+4.226
Vdc
250
AUTO
10
+4.201
to
+4.226
Vdc
251
AUTO
11
+4.201
to
+4.226
Vdc
236
CIP
RUN/HOLD/OFF switch to HOLD.
252
AUTO
12
+4.201
to
+4.226
Vdc
237
CIP
COUNTER INHIBIT(SEC) dial to 16.01.
253
AUTO
13
+4.201
to
+4.226
Vdc
238
CIP
RUN/HOLD/OFF switch to RUN.
254
AUTO
14
+4.201
to
+4.226
Vdc
239
CIP
Observe INHIB lamp illuminated.
255
AUTO 15
+4.201 to +4.226 Vdc
240
CIP
RUN/HOLD/OFF switch to HOLD.
256
AUTO 16
+4.201 to +4.226 Vdc
241
CIP
COUNTER INHIBIT(SEC) dial to 16.03.
Corrective action for steps 244 through 256 is:
242
CIP
RUN/HOLD/OFF switch to RUN and observe the waveform drops to zero volts for 315 to 585 msec, then
(30)
expands to a level of 0.2 VP-P.
Replace card Al. Check A1 pin 62.
257
AUTO 3
+9.671 to +9.729 Vdc
243
PPE
Set SEQUENCE switch to the positions indicated and observe the DMM reading for each switch setting.
(31)
<*> (See Step 174)
258
AUTO 4
+3.947 to +3.971 Vdc
(32)
259
AUTO 7
+10.24 to +10.34 Vdc
(33)
3-85