Table 3-11. DMS-D Performance Test-Continued
Table 3-11. DMS-D Performance Test-Continued
STEP
UNIT
PROCEDURE CORRECTIVE ACTION
STEP
UNIT
PROCEDURE
CORRECTIVE ACTION
Replace card A5. Check A5 pins
684
Compute the difference between step 681 and 682. The difference
40, 103, and 111.
shall be 0.1113 to 0.1207 Vdc.
674
PPE
SEQUENCE switch to AUTO 4.
A
(91)
675
CIP
COUNTER INHIBIT(SEC) dial to 00.33.
676
PPE
Momentarily actuate TEST START switch.
685
Compute the difference between step 681 and 683. The difference
677
ITP
Set INTERNAL TEST switch to the positions indicated. Then set
shall be 0.1113 to 0.1207 Vdc.
CIP
RUN/HOLD/OFF switch to HOLD, the COUNTER INHIBIT(SEC) dial to
Replace card A5. Check A5 pins
DMM
the positions indicated, the RUN/HOLD/OFF switch to RUN and
40, 103 and 111.
observe the DMM reading for each series of switch settings.
Record the DMM reading in steps 681, 682, and 683 for use
686
PPE
SEQUENCE switch to AUTO 5.
in steps 684 and 685.
687
CIP
RUN/HOLD/OFF switch to OFF.
688
ITP
INTERNAL TEST switch to 1.
689
DMM
Observe a reading of -0.8780 to -0.8860 Vdc. Record for use in
INTERNAL TEST
COUNTER INHIBIT(SEC)
Reading
steps 692 and 695.
678
1
00.34
-0.0050 to +0.0050 Vdc
(94)
(92)
679
1
00.35
+0.5220 to +0.5280 Vc
d
690
ITP
INTERNAL TEST switch to 2.
(87)
691
DMM
Record the reading for use in step 692.
680
00.99
+0.5220 to +0.5280 Vdc
692
Compute the difference between step 689 and 691. The difference
(87)
shall be 0.1113 to 0.1207 Vdc.
(91)
681
01.01
+0.8780 to +0.8860 Vdc
(93)
Record reading.
693
ITP
INTERNAL TEST switch to 3.
694
DMM
Record the reading for use in step 695.
682
2
Record reading.
695
Compute the difference between step 689 and 694. The difference
683
3
Record reading
shall be 0.1113 to 0.1207 Vdc.
3-101