TM9-4935-481-14-1
C3
Table 3-11. DMS-D Performance Test-Continued
Table 3-11. DMS-D Performance Test-Continued
STEP
UNIT
PROCEDURE
CORRECTIVE ACTION
STEP
UNIT
PROCEDURE
CORRECTIVE ACTION
749
Compute the difference between step 743 and 748. The difference shall be 0.1113 to 0.1207 Vdc.
INTERNAL TEST
COUNTER INHIBIT(SEC)
Reading
Replace card A5. Check A5 pins
40, 103, and 111.
736
2
Record reading.
750
CIP
RUN/HOLD/OFF switch to RUN.
737
3
Record reading.
751
PPE
SEQUENCE switch to AUTO 12.
752
PPE
Momentarily actuate TEST START switch.
738
Compute the difference between step 735 and 736. The difference
753
ITP
Set INTERNAL TEST switch to the positions indicated. Then set
shall be 0.1113 to 0.1207 Vdc.
CIP
DMM
RUN/HOLD/OFF switch to HOLD, the COUNTER INHIBIT(SEC) dial to
the positions indicated, the RUN/HOLD/OFF switch to RUN and wait
739
Compute the difference between step 735 and 737. The difference
for the INHIBIT lamp to illuminate, then observe the DMM reading
shall be 0.1113 to 0.1207 Vdc.
for each series of switch settings.
Replace card A5. Check A5 pins
754
Record the DMM reading in steps 758, 759 and 760 for use in
40, 103, and 111.
steps 761 and 762.
740
PPE
SEQUENCE switch to AUTO 11.
INTERNAL TEST
COUNTER INHIBIT(SEC)
Reading
741
CIP
RUN/HOLD/OFF switch to OFF.
755
1
16.99
+3.403 to +3.437 Vdc
742
ITP
INTERNAL TEST switch to 1.
(98)
743
DMM
Observe a reading of +6.151 to +6.213 Vdc. Record for use in steps 746and 749.
756
17.01
+3.964 to +4.004 Vdc
744
ITP
INTERNAL TEST switch to 2.
(101)
745
DMM
Record the reading for use in step 746.
746
Compute the difference between step 743 and 745. The difference shall be 0.1113 to 0.1207 Vdc
757
18.39
+3.964 to +4.004 Vdc
(101)
747
ITP
INTERNAL TEST switch to 3.
758
18.41
+3.403 to +3437 Vdc
.
748
DMM
Record the reading for use in step 749.
(98)
Record reading.
759
2
Record reading.
760
3
Record reading.
3-104