Table 3-11. DMS-D Performance Test-Continued
Table 3-11. DMS-D Performance Test-Continued
STEP
UNIT
PROCEDURE
CORRECTIVE ACTION
STEP
UNIT
PROCEDURE
CORRECTIVE ACTION
1671
TP
Momentarily actuate TEST START switch.
1672
CIP
Observe INHIB lamp illuminated before continuing.
1650
VWS
Adjust AMPTD(VOLTS) control to +0.324 0.0002 Vdc.
8
1673
TP
Observe GO lamp extin uished and NO GO lamp illuminated.
g
DMM
1674
CIP
COUNTER INHIBIT(SEC) dial to 01.91.
1651
TP
Momentarily actuate TEST START switch.
1675
MCP
DMM switch to DMM INPUT.
1652
CIP
Observe INHIB lamp illuminated before continuing.
1676
VWS
Adjust AMPTD(VOLTS) control to +0.992 0.0020 Vdc.
0
1653
TP
Observe GO lamp extinguished and NO GO lamp illuminated.
DMM
1654
CIP
COUNTER INHIBIT(SEC) dial to 00.21.
1677
MCP
DMM INPUT switch to SEL A.
1655
VWS
+DC/PULSE/-DC switch to PULSE.
1678
MCP
Adjust COMPARATOR THRESHOLD B control to -1.012 0.0010 Vdc.
1656
VWS
Adjust AMPTD(VOLTS) control to +7.50 0.0050 Vdc.
0
DMM
DMM
1679
TP
Momentarily actuate TEST START switch.
1657
MCP
DMM INPUT switch to SEL A.
1680
CIP
Observe INHIB lamp illuminated before continuing.
1658
MCP
Adjust COMPARATOR THRESHOLD B control to -7.89 0.0010 Vdc.
0
1681
TP
Observe GO and NO GO lamps illuminated.
DMM
1682
MCP
Adjust COMPARATOR THRESHOLD B control to -1.032 0.0010 Vdc.
1659
TP
Momentarily actuate TEST START switch.
DMM
1660
CIP
Observe INHIB lamp illuminated before continuing.
1683
TP
Momentarily actuate TEST START switch.
1661
TP
Observe GO lamp illuminated and NO GO lamp extinguished.
1684
CIP
Observe INHIB lamp illuminated before continuing.
1662
MCP
Adjust COMPARATOR THRESHOLD B control to -8.040 0.001 Vdc.
1685
TP
Observe GO lamp extinguished and NO GO lamp illuminated.
DMM
1686
MCP
Adjust COMPARATOR THRESHOLD B control to -0.9720 0.0010 Vdc.
1663
TP
Momentarily actuate TEST START switch.
DMM
1664
CIP
Observe INHIB lamp illuminated before continuing.
1687
TP
Momentarily actuate TEST START switch.
1665
TP
Observe GO lamp extinguished and NO GO lamp illuminated.
1688
CIP
Observe INHIB lamp illuminated before continuing.
1666
MCP
Adjust COMPARATOR THRES OLD B control to -7.110 0.0010 Vdc.
H
1689
TP
Observe GO and NO GO lamps illuminated.
DMM
1690
MCP
Adjust COMPARATOR THRESHOLD B control to -0.9520 0.0010 Vdc.
1667
TP
Momentarily actuate TEST START switch.
DMM
1668
CIP
Observe INHIB lamp illuminated before continuing.
1691
TP
Momentarily a tuate TEST START switch.
c
1669
TP
Observe GO lamp illuminated and NO GO lamp extinguished.
1670
MCP
Adjust COMPARATOR THRESHOLD B control to -6.960 0.0010 Vdc.
DMM
3-135